Skz1044A Manual Operation K Na Digital Flame Photometer Flame Spectrometer Analyzer Digital Flame Photometer Instrument
Overview Package Size 40.00cm * 25.00cm * 50.00cm Package Gross Weight 8.000kg Lead Time 14 days (1 - 1 Pieces) To be ne;
Overview
To be negotiated ( > 1 Pieces)
Detailed Photos
Basic Info.
Model NO. | SKZ1044A |
Mass Scope of Workpiece | <900kg |
Type | Material Testing Machine |
Maxcapacity | <1000KN |
Accuracy Grade | 1 |
Load Way | Mechanical Load |
Loading Method | Static Load |
Display | Digital |
Control | Manual Control |
Weight | 0-100Kg |
Power Source | AC220V |
Sample Volume | <6ml / Min |
Response Time | <8s |
Photoelectric Conversion Elements | Silicon Photovoltaic Cells |
Certificate | CE |
Spectroscopic Methods | Interference Filter |
Items | Specification |
Test Range | K: 0 ~ 100ug / Ml Na: 0 ~ 160ug / Ml |
Transport Package | Wooden Box |
Specification | 40*25*50 cm 8kg |
Trademark | SKZ |
Origin | China |
HS Code | 9027809990 |
Production Capacity | 500 |
Packaging & Delivery
Package Size 40.00cm * 25.00cm * 50.00cm Package Gross Weight 8.000kg Lead Time 14 days (1 - 1 Pieces)To be negotiated ( > 1 Pieces)
Product Description
Product DescriptionSKZ1044A Digital Flame Photometer
Flame photometer is a relative measurement of the instrument, the concentration value of the sample under test conditions in the same relative value of the concentration of the standard solution. So before the test, you must first prepare a set of corresponding standard solution. Then the calibration operation, the manual or the standard curve by the instrument, the last to be tested on the test sample, to obtain the concentration value or other datarequired for calculation.Applications1) cement, glass, ceramics, refractories and other materials of the test;2) fertilizers, soil of the test;3) mining, petroleum, metallurgy, chemical product testing;4) pharmaceuticals, beverages and other food testing;5) municipal solid waste test;6).scientific research, health, education and other fields all kinds of experimental tests.PrincipleFlame Photometer is emission spectroscopy, It utilizes the heat flame itself, stimulate the part of the alkaline earth metalatoms, so that these atoms absorb energy from a first energy level transition, when it falls to a normal energy level, it isnecessary to release energy, the energy released have spectral characteristics, that is, in a certain wavelength range.For example, the salt in a fire, flame yellow, because sodium atoms in the flame down to a normal level when the energy spectrum of the energy released is yellow. People often call this the flame reaction. Different alkali metal or alkaline earth metal in the flame of different colors, each with a different filter, it can be qualitatively tested. The flame intensity is directlyproportional to the solubility atoms contained in the solution, which constitutes the basis for a quantitative test. This method is usually called flame photometry, such instruments are usually referred to as flame photometer.SpecificationTechnical parameters:
Items | Specification | |
Test range | K: 0 ~ 100ug / ml | |
Na: 0 ~ 160ug / ml | ||
Stability | Continuous injection of standard solution, the relative maximum change of indication within 15s ≤ 3% | |
Measured once per minute, a total of 6 measurements, the relative maximum indication of changes in equipment ≤ 15% | ||
Repeatability | ≤ 3% | |
Linearity error | K: ≤0.005mmol/L (0.0100 ~ 0.0800) mmol/L | |
Na: ≤0.03mmol/L (0.0500 ~ 0.400) mmol/L | ||
Detection limit | K: ≤ 0.004mmol/L | |
Na: ≤ 0.008mmol/L | ||
Filter | K filter | absolute value of peak wavelength error≤ 7nm |
Half width≤15nm | ||
background transmittance≤ 0.5% | ||
Na filter | absolute value of peak wavelength error≤ 5nm | |
half-width≤ 15nm | ||
background transmittance≤ 0.5% | ||
Response time | <8s | |
Sample volume | <6mL / min |