Overview Package Size 50.00cm * 60.00cm * 80.00cm Package Gross Weight 20.000kg Automatic Potentiometric Titrator EC-GT7;
Overview
Imported corrosion-resistant PTFE rotary valve, PTFE corrosion-resistant burette, high-power mixing table, coil for magnetic stirring
High closed-loop control ensures titration
The resolution of burette is 1uL
The error of high-precision burette is 0.003mm
Automatic check of instrument status upon startup
The overall pipeline is fully sealed, making replacement quick and convenient
Real time display of titration curves and measurement results for easy operation
Supports various titration methods such as acid-base titration, redox titration, precipitation titration, complexation titration, and non-aqueous titration
Complies with GMP specifications, titration results meet GLP requirements
Automatic sample loading, automatic cleaning of pipelines, automatic measurement of experimental results, automatic recording and plotting of real-time changes in pH titration volume during titration.
Complies with the 2015 Pharmacopoeia
Equipped with comprehensive multi-level user permission management functions, achieving audit tracking and permission management functions
Application field:
Acid-base titration, such as oxalic acid (aqueous/non aqueous) content
Oxidation-reduction titration, such as iodine value, iron (II) content
Precipitation titration, such as chloride content
Complexometric titration, such as calcium content
Titration with polarized electrodes (Ipol, Upol)
Application
Basic Info.
Model NO. | EC-GT70 |
Supply Area | Global |
Function | Potentiometric Titrator |
Lead Time | 7-15working Days |
Transport Package | Wooden, Carton |
Specification | EC-GT70 |
Trademark | SRI |
Origin | China Mainland |
HS Code | 8419899090 |
Production Capacity | 500 Sets Per Year |
Packaging & Delivery
Package Size 50.00cm * 60.00cm * 80.00cm Package Gross Weight 20.000kgProduct Description
Automatic Potentiometric Titrator EC-GT70
Imported corrosion-resistant PTFE rotary valve, PTFE corrosion-resistant burette, high-power mixing table, coil for magnetic stirring
High closed-loop control ensures titration
The resolution of burette is 1uL
The error of high-precision burette is 0.003mm
Automatic check of instrument status upon startup
The overall pipeline is fully sealed, making replacement quick and convenient
Real time display of titration curves and measurement results for easy operation
Supports various titration methods such as acid-base titration, redox titration, precipitation titration, complexation titration, and non-aqueous titration
Complies with GMP specifications, titration results meet GLP requirements
Automatic sample loading, automatic cleaning of pipelines, automatic measurement of experimental results, automatic recording and plotting of real-time changes in pH titration volume during titration.
Complies with the 2015 Pharmacopoeia
Equipped with comprehensive multi-level user permission management functions, achieving audit tracking and permission management functions
Application field:
Acid-base titration, such as oxalic acid (aqueous/non aqueous) content
Oxidation-reduction titration, such as iodine value, iron (II) content
Precipitation titration, such as chloride content
Complexometric titration, such as calcium content
Titration with polarized electrodes (Ipol, Upol)
Measuring range | -2000.0MV~+2000.0MV; -20.000PH~+20.000PH |
Resolving power | 0.1MV 0.001PH |
Accuracy | 0.1MV 0.003PH |
Measurement mode | MEAS: Measure pH/mV/T CAL: pH correction (multi point) SET set endpoint mode MET equivalent titration DET dynamic titration |
Titration mode | Acid-base titration, redox titration, precipitation titration, complexation titration, non aqueous titration |
Temperature measurement range | -5~120ºC |
Temperature measurement accuracy | ±0.1ºC |
Communication method | 2 USB, 232 serial ports, Ethernet |
Filling unit | Built-in liquid dispensing unit |
Simultaneously adding liquid module unit | (Standard configuration) Includes 2 sets of built-in titration dosing channels |
Number of expandable dispensing channels | 8 channel |
Burette specification | 1ml 5ml 10ml 25ml 50ml(10ml standard) |
Burette resolution | 1/48,000 |
Burette rehydration time | 10 seconds (100% filling speed) |
Control unit | PC includes software (PC comes separately) |
Automatic sampler | Can be used in conjunction with GT robot sampling system for analyzing large batches of samples |
Self inspection function | Automatic check of instrument status upon startup |
Calibration function | Yes |
Application editing function | Yes |
Storage space | Greater than 300G (greater than 10000 pieces) |
Preset method | Greater than 300G (greater than 10000 pieces) |
Compliant with FDA, GMP, GLP | Audit tracking, hierarchical management, electronic signature |
Piping | PTFE titration pipeline (corrosion resistant) |
Mixing method | Top mounted spiral stirrer or magnetic stirring |
PTFE valve | Imported corrosion-resistant rotary valve |
Micro detection electrode | Imported type |
Microinjection | Imported type |
Determination of acid value in food - Determination of chloride ions in cement
Determination of peroxide value in food - Determination of mercaptan sulfur
Determination of reducing sugars in food - Determination of electrolyte
Determination of fatty acid value in corn flour Determination of surfactants
Free fatty acid determination COD determination in wastewater
Determination of succinic acid content in traditional Chinese medicine Pinellia ternata Determination of free acid/base
Determination of organic matter in soil.